Pacemakers: why more older patients need one

In this article, Dr Chris Plummer explains that many older people that would benefit from pacemakers are not being identified.

Pacemakers are the only safe treatment for bradycardia. We now take this remarkable technology for granted. We have extremely reliable, safe, cost-effective devices with an enormous evidence base of improving symptoms and prognosis. Yet pacing is greatly underused in the UK, with implant rates far below those in our neighbouring European countries2. Because the need for pacing is so closely associated with age, this is an important elderly care issue which must be addressed if we are to give our patients the standard of care they deserve.

Arne Larsson was born on 26 May 1915. At the age of 43 he was diagnosed with a fatal disease – complete heart block with Stokes-Adams attacks. His wife would not accept her husband’s inevitable death and, having heard of animal experiments with cardiac pacing, she persuaded a surgeon, Åke Senning from the Karolinska Institute, and an engineer, Rune Elmqvist, who worked for a medical electronics firm in Stockholm, to build (in Elmqvist’s kitchen) and then implant, the first human pacemaker on 8 October 19581.

Arne Larsson died at the age of 86 on 28 December 2001, 43 years, 22 pacemakers and five electrode systems later, from an unrelated malignancy.

Background to pacemakers

Contraction of the heart, like all muscle, is coordinated by electrical impulses. Unlike skeletal muscle, however, the heart receives only tonic neural input which comes from sympathetic and parasympathetic nerves. The precise co-ordination of myocardial contraction is controlled by specialised muscle cells in the walls of the heart itself.

Impulse formation

The heart rate is controlled by the least electrically stable tissue in the heart. This is normally the sinoatrial (SA) node, a cluster of specialised myocytes high in the right atrium. Intracellular recordings have shown spontaneous depolarisation during diastole (phase 4 depolarisation) in these pacemaker cells controlled by the If current [Fig f current3. The slope of this depolarisation, and thus the heart rate, is increased by adrenergic stimulation from sympathetic nerves or circulating catecholamines, and decreased by acetylcholine from vagal input.

Impulse conduction

From the SA node, impulses travel through the atrial myocardium, resulting in atrial systole expelling blood into the ventricles. When an impulse reaches the atrio-ventricular (AV) node at the top of the interventricular septum it is delayed by 120 to 200ms (the normal PR interval) allowing time for ventricular filling before the impulse is conducted to the His bundle, down the bundle branches and into the Purkinje system.

Because these specialised conduction tissues conduct impulses at high velocity through gap junctions, this results in rapid ventricular contraction, a narrow QRS complex (≤120ms) on the surface electrocardiogram (ECG) and an efficient co-ordinated ventricular contraction [Fig 4]. Another important property of the AV node is that as it is presented with increasingly rapid inputs, these are increasingly delayed (decremental conduction) until, at the Wenckebach point, an increasing proportion are blocked from reaching the ventricles. This prevents the very rapid ventricular stimulation, which would otherwise occur in atrial fibrillation (AF) from inducing ventricular fibrillation.

Failure of impulse formation

If spontaneous SA node depolarisation slows inappropriately or fails completely, another focus within the atria – the atrio-ventricular (AV) node – the His bundle, one of the bundle branches or the ventricular myocardium itself will take over this intrinsic pacemaker function in a series of ‘fail-safe’ mechanisms. This results in bradycardia at rest or the failure of heart rate to increase appropriately on exercise – sinus node dysfunction (SND).

SND is associated with fibrosis and loss of pacemaker cells. This process is associated with age, but the cause is often unknown with only occasional evidence of direct infiltration with malignancy, inflammatory disease, cardiomyopathy, ischaemic or direct surgical damage.

The rate of SA node depolarisation is reduced by high vagal tone, as seen in highly trained athletes, and neuro-cardiogenic reflex syncope. Other medical conditions such as hypothyroidism, elevated intracranial pressure, obstructive jaundice and hypothermia depress sinus node function as do beta-blockers (ß-blockers), rate-limiting calcium antagonists, lithium and the new If blocker, f blocker, ivabradine.

Failure of impulse conduction

Delayed conduction through the AV node (shown by a PR interval of >200ms) represents first degree heart block. While this is often asymptomatic, it can reduce cardiac output as there is often mitral and tricuspid regurgitation before ventricular contraction. In second degree heart block, not all impulses are conducted to the ventricles. In type 1 there is a progressive lengthening of the PR interval before a non-conducted P wave (Wenckebach) while in the type 2 form, there is a fixed PR interval in the conducted beats with non-conducted beats usually in a 2:1 or 3:1 ratio. Complete failure of AV conduction is known as third degree heart block. Depending on the escape rhythm, this can be relatively asymptomatic or result in syncope or even sudden cardiac death.

Failure of conduction in one or more the bundle branches results in the delayed contraction of the ventricular wall it supplies as the impulse has to spread through normal myocardium rather than the specialised conduction tissue. This results in the broad QRS complex seen on the surface ECG. In people with normal ventricular function, this is usually asymptomatic, but in those with impaired function, any loss of co-ordination between the medial (septal) and lateral walls of the left ventricle can cause or worsen heart failure.

Conduction tissue fibrosis is the most common cause of heart block in older adults. It can also be associated with aortic valve stenosis or ischaemic damage. Drugs that affect AV node conduction include ß-blockers, rate-limiting calcium antagonists and digoxin, which acts predominantly by augmenting vagal tone.

Pacemakers

Pacemaker technology has evolved rapidly from its beginnings in the 1950s. Devices have become smaller and increasingly sophisticated [Fig 5]. Functionally, all pacing systems contain one or more leads with electrodes in the heart, connected to battery powered circuitry which senses intrinsic impulses, time intervals between these impulses and can deliver electric pulses to the heart to trigger contraction. Under local anaesthetic, leads are implanted in the heart chambers, usually via the subclavian vein under x-ray guidance. When correctly positioned radiographically and electrically, they are attached to the pacemaker pulse generator, which is placed in a ‘pocket’ fashioned immediately in front of – or occasionally beneath – the pectoral muscle.

The patient is discharged the following day after satisfactory chest x-ray and pacing checks performed by radio telemetry.

Patients are usually followed up at six weeks, six months and then annually until the battery is depleted, usually after seven to 10 years, when they are electively readmitted as a day case for generator replacement.

Pacemaker nomenclature

Pacemakers are described by a code developed jointly by the UK and US pacing societies4 . This comprises three initial letters: the first refers to the chamber paced, the second to the chamber where sensing of impulses takes place and the third to the device’s response to sensing. The devices most commonly used in the UK are VVI (ventricular pacing, ventricular sensing and inhibition of pacing by sensing), AAI (atrial pacing, atrial sensing and inhibition of pacing by sensing) and DDD (dual chamber pacing, dual chamber sensing and a dual response of inhibition or triggering of ventricular pacing at a programmed interval after atrial depolarisation). An optional fourth letter R refers to rate response, indicating that the device automatically increases the pacing rate on exercise.

Indications for implantation

Pacemaker implantation is indicated in all those with symptomatic bradycardia without a reversible cause – there is no other safe treatment. The management of those without symptoms is based on epidemiological data and the results of clinical trials. The common internationally accepted pacing indications are summarised below.

Pacing for symptoms

Although sinus node dysfunction is the most common reason (26.7 per cent) for pacemaker implantation in the UK5, there is little evidence this improves survival – the indication is to improve symptoms. Because these can be relatively nonspecific, it is important to correlate symptoms with sinus bradycardia or the failure of heart rate to increase on exercise (chronotropic incompetence), usually by prolonged ECG (Holter) monitoring. In the absence of this correlation, pacemaker implantation can be justified if sinus bradycardia at

Pacing for prognosis

There are conditions where there is good clinical evidence life expectancy is increased by pacing even in the absence of symptoms. These include high degree (third- and second- degree type 2) AV block, those with evidence of widespread conduction system disease (such as alternating left and right bundle-branch block) and those with lesser degrees of AV block (second degree type 1, bi- or tri-fascicular block) where block in or below the His bundle has been demonstrated on electrophysiology study. Further information on pacemaker indications is available in the international guidelines7.

Mode of pacing

The first pacemakers delivered single chamber ventricular pacing only. The development of dual chamber pacing in 19638 allowed resynchronisation of atria and ventricles in complete heart block. Although this was clearly an important advance in pacing technology, in most bradycardia patients the magnitude of benefit derived from pacing is far larger than any difference between pacing modes. Evidence from randomised trials suggests that ventricular pacing appears to increase the risk of developing atrial fi brillation9,10,11,12 and symptomatic heart failure12 in patients with SND.

In patients paced for AV block, the Pacemaker Selection in the Elderly (PASE) trial13 and Canadian Trial of Physiologic Pacing (CTOPP)14 failed to demonstrate any significant difference in mortality or stroke between pacing modes, although the UK Pacing and Cardiovascular Events (UKPACE)15 study showed ventricular pacing (VVI) was associated with a higher risk of stroke, transient ischemic attack or other thromboembolism (2.5 per cent per year) compared to dual chamber (P=0.04) pacing, which was not different from ventricular rate-responsive pacing (1.7 per cent per year, P=0.93).

In February 2006, the UK National Institute for Health and Clinical Excellence (NICE) issued guidance on pacing mode in symptomatic bradycardia16. They recommended dual chamber pacing, except in:

  • patients with SND and normal AV conduction when single chamber atrial pacing is appropriate;
  • in those with continuous atrial fibrillation when single chamber ventricular pacing is indicated; and
  • in those where there are patient-specific factors, such as frailty or the presence of co-morbidities, when the balance of risks and benefits is in favour of single chamber ventricular pacing.

New pacemakers indications

Although pacemakers were developed as a treatment for bradycardia, advances in computer technology over the past 40 years have allowed the development of other pacing indications. One of the first was the termination of supra-ventricular tachycardias by high rate atrial pacing. This is now only rarely necessary as most of these tachycardias are cured by catheter ablation, but the same technology is used in implantable cardioverter defibrillators (ICDs) to pace terminate ventricular tachycardia without the need for defibrillation. Many current pacemakers can now store ECGs and intra-cardiac electrograms, allowing diagnosis and then monitoring of the treatment of arrhythmias. With the recognition that atrial fibrillation is strongly associated with SND and that individual paroxysms of AF can be triggered by sinus pauses, algorithms have been developed which suppress these triggers by maintaining a high proportion of atrial pacing.

One of the most interesting developments in pacing over the past decade has been pacing for heart failure. Many heart failure patients have atrio-ventricular and intra-ventricular conduction delays, especially left bundle branch block, and these can worsen ventricular function by desynchronising contraction. Cardiac resynchronisation devices have been developed to pace the left ventricular free wall via the coronary sinus, and the interventricular septum via the right ventricle. Clinical trials have shown improvements in acute haemodynamic measurements, patient symptoms and, most recently, mortality17.

Where is the block?

Pacemaker implantation improves both quality and quantity of life in appropriate patients, it is highly cost effective, and has recently been endorsed by NICE. Over the past decade the number of devices implanted has increased exponentially around the world, particularly in North America and many European countries, but the UK lags behind most developed countries. We do not yet know the exact sites or causes of this service block, although it appears to be present at all levels, from patients not reporting their symptoms, to under investigation in primary and secondary care.

It is clear the majority of pacing indications occur in the elderly and that no one should be excluded from pacing for symptomatic bradycardia on grounds of age or cognitive function. We must all consider bradycardia as a cause of our patients’ often non-specific symptoms and investigate them appropriately if we are to give them the standard of care they deserve and increase rate of pacemaker implantation in the UK to European standards.


Dr Chris Plummer, Consultant Cardiologist, Newcastle

Conflict of interest: Dr Plummer has received educational and travel grants from the pacemaker manufacturers Guidant, St Jude Medical and Medtronic


References

  1. Elmqvist R, Senning A. Implantable pacemaker for the heart. In: Smyth CN, ed. Medical electronics: proceedings of the Second International Conference on Medical Electronics, Paris, June 24–27, 1959. London: Iliffe, 1960:253-4. abstract 2
  2. Mond HG, Irwin M, Morillo C, Ector H. The world survey of cardiac pacing and cardiovert defibrillators: Calendar Year 2001. PACE 2004; 27:955- 964
  3. Brown HF, DiFrancesco D, Noble SJ. How does adrenaline accelerate the heart? Nature 1979; 280:235-236
  4. Bernstein AD, Parsonnet V. The NASPE/BPEG pacemakerlead code (NBL code) Pacing Clin Electrophysiol 1996; 19:1535-6
  5. Cunningham D, Rickards A, Cunningham M. National Pacemaker Database, United Kingdom and Republic of Ireland, National ICD Database annual report 2002
  6. Kenny RA, Richardson DA, Steen N, et al. Carotid sinus syndrome: a modifiable risk factor for non-accidental falls in older adults (SAFE PACE). J Am Coll Cardiol J Am Coll Cardiol 2001; 38:1491-6
  7. Gregoratos G, Abrams J, Epstein AE, et al. ACC/AHA/ NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/ AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation 2002; 106:2145-61
  8. Nathan DA, Center S, Wu C-Y, Keller JW. An implantable synchronous pacemaker for the long term correction of compete heart block. Circulation 1963; 27:682-5
  9. Andersen HR, Thuesen L, Bagger JP, et al. Prospective randomised trial of atrial versus ventricular pacing in sick-sinus syndrome. Lancet 1994; 344:1523-8
  10. Andersen HR, Nielsen JC, Thomsen PEB, et al. Longterm follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome. Lancet 1997; 350:1210-6
  11. Nielsen JC, Andersen HR, Thomsen PEB, et al. Heart failure and echocardiographic changes during long-term follow-up of patients with sick sinus syndrome randomized to single chamber atrial or ventricular pacing. Circulation 1998; 97:987-95
  12. Lamas GA, Lee KL, Sweeney MO, et al. Ventricular pacing or dual chamber pacing for sinus node dysfunction (MOST). N Engl J Med N Engl J Med 2002; 346:1854-62
  13. Lamas GA, Orav EJ, Stambler BS, et al. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared to with dual chamber pacing. Pacemaker Selection in the Elderly Investigators. N Eng J Med 1998; 338:1097-1104
  14. Connolly SJ, Kerr CR, Gent M et al. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. Canadian Trial of Physiologic Pacing Investigators. N Engl J Med 2002;342:1385-91
  15. Toff WD, Camm AJ, Skehan JD. For the United Kingdom Pacing and Cardiovascular Events (UKPACE) Trial Investigators. Single-Chamber versus Dual-Chamber Pacing for High-Grade Atrioventricular Block. N Engl J Med N Engl J Med 2005; 353:145-55
  16. Dual-chamber pacemakers for symptomatic bradycardia: Guidance. National Institute for Clinical Excellence 23 February 2005 Technology appraisal 88
  17. Cleland JGF, Daubert J-C, Erdmann E, et al. The Effect of Cardiac Resynchronization on Morbidity and Mortality in Heart Failure. N Engl J Med 2005; 352:1539-49